ELL-1005 高速大电流电子负载

厂  牌 : KEISOKU GIKEN

Load section

Max rating
Current405A
Voltage30V
Internal min resistance1.5m ohm
Internal inductance (Typ)80nH
Min operating voltage(*6)0.6V(405A) / 0.3V(202.5A) / 0.1V(67.5A)
Peak power(*5)12000W (20us or less)
1290W(20s or less)
Rated power1000W
CC mode
Current range45A405A
Setting range0 - 45A0 - 405A
Nominal resolution(*1)3mA30mA
Accuracy+/- 0.2% of stg. +/- 60mA+/- 0.2% of stg. +/- 500mA
CR mode
Voltage range4V
Current range45A405A
Resistance range (Setting range)0.0167 - 166.667 ohm1.7m - 16.667 ohm
Nominal resolution(*1)6ms60ms
Accuracy(*2)+/- 0.5% of Conv. Curr. +/- 0.2% of f.s.
 
Voltage range30V
Current range45A405A
Resistance range (Setting range)0.05 - 500 ohm0.005 - 50 ohm
Nominal resolution(*1)6ms60ms
Accuracy(*2)+/- 0.5% of Conv. Curr. +/- 0.2% of f.s.
CV mode
Voltage range4V30V
Setting range0 - 4V0 - 30V
Nominal resolution(*1)0.5mV5mV
Accuracy+/- 0.2% of stg. +/- 0.2% of f.s.
Response timeFast/Slow (300us/10ms)
CP mode
Current range45A405A
Power range (Setting range)0 - 120W0 - 1000W
Nominal resolution(*1)15mW75mW
Accuracy+/- 2.5% of stg. +/- 1% of f.s.
External control mode 
Current range45A405A 
Setting range0 - 45A0 - 405A 
Nominal resolution(*1)6mA60mA 
Accuracy(*4)+/- 0.2% of stg. +/- 0.5% of f.s.
Control voltage0 - 10V 
Dynamic load mode 
Dynamic mode 
Current range4A405A 
Control methodSwitching 
Loading modeCC mode 
Setting period- 20ms / - 200ms / - 2s / - 20s / - 60s 
Period resolution1us / 10us / 100us / 1ms / 10ms 
Operating modeContinuous, Single 
Slew rate0.05 - 5A/us0.5 - 50A/us 
Min load response time(*3)500ns 
Sequence mode 
Loading modeCC / CR / CV / CP mode 
Max steps1024 
Step duration1ms - 10min 
Suration resolution1ms(1ms - 100ms) / 100ms(100ms - 10min) 
Short mode 
Short current (MAX)405A 
Current limit setting 
Current range45A405A 
Setting range0.3 - 45A3 - 405A 
Nominal resolution(*1)300mA3A 
Parallel operation 
Master-Slave operation. One master unit can control up to 4 units of slaves of same voltage rating. For example: If the master unit is ELA-305 then ELA-155, ELA-305 or ELA-1005 may be connected as slave unit. 

*1 : Nominal resolution: Approximated resolution of the range. / *2 : Conv.Curr: Ideal current value being calculated by [input / set R] / *3 : Min load response time: Minimum time required by the load when setting slew rate. / *4 : Rin: ELA-155 : 100kohm(typ) ELB-155 : 200kohm(typ) ELC-155 : 1500kohm(typ) / *5 : External controlsetting accuracy: Setting accuracy is defined when external control voltage of 10V is fed. / *6 Setting accuracy in CR mode is valid when applied 10% or more voltage of the voltage range. / *7 : It may vary in accordance to the internal temperature. / *8 Minimum operating voltage vary according to the current value. / *9 At rear loading terminals.

Mesurement section
DC voltage
Range4.0000V30.000V
Resolution0.1mV1.0mV
Accuracy(*1)+/- 0.05% of rdg. +/- 0.05% of f.s.
Measuring time(*2)Approx. 100ms
DC current
Range(*3)45.0000A405.000A
Resolution0.3mA3mA
Accuracy(*1)+/- 0.2% of rdg. +/- 0.2% of f.s.
Measuring time(*2)Approx. 100ms
Power
Measuring method(*4)By calculation [Voltage x Current]
Measuring time(*2)Approx. 200ms

*1 Guaranteed for 6 months at 23 +/- 5C, <70% RH / *2 Measurement time when set at same measuring mode and range. / *3 Current measurement range depends on load setting range. / *※4 Absolute value of the measurement.

General
General
Loading terminalsOn front panel
Rates AC inputAC100 - 240V +/- 10% 50/60Hz
Power consumption70VA or less (when in AC100V)
Operating tempperature5 - 40C
Cooling systemForced cooling fan
Dimentions430(W) X 128.6(H) X 420(D) excluding projected part
WeightApprox. 16Kg
Interface
GPIBComforming to IEEE488.1
USBComforming to USB1.1
External control voltage0 - 10V
Protection, Alarm function
Over current protectionProtect load section by limitting the current
Over power protectionProtect load section by limitting the power
Over voltage alarmGenerate alarm when over voltage is applied
Temperature ProtectorGenerate alarm and cut off the output when overheated
Reverse connection alarmGenerate alarm when reverse voltage is applied
  • Noted that the alarm function does not protect the instrument. Remove the cause of alarm to protect the instrument.
TRIG OUT
Output(*1)Photo coupler output
Output voltageCC1+5V (Typ)
CC20V (Typ)
CURRENT MONITOR
Output(*2)1V / 405A f.s.
Output resistance50 ohm
Accuracy(*3)+/- 1% of Conv.Volt. +/- 1% of f.s.

*1 TRIG OUT: Isolated / *2 CURRENT MONITOR: Not isolated / *3 Full scale of the current monitor and rated current are different in current range High, Mid and Low.

 